Flowline LI90 Commander Manual de usuario

Busca en linea o descarga Manual de usuario para Herramientas Flowline LI90 Commander. Flowline LI90 Commander User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 57
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
27 AUG 08 PRELIMINARY MANUAL 1 of 57
Rev 1_3 Commander LI90 Series
Commander
LI90 Series
Controller
User Manual
Flowline Inc.
10500 Humbolt Street
Los Alamitos, CA 90720
Tel: (562) 598-3015
Fax: (562) 431-8507
www.flowline.com
Vista de pagina 0
1 2 3 4 5 6 ... 56 57

Indice de contenidos

Pagina 1 - Controller

27 AUG 08 PRELIMINARY MANUAL 1 of 57 Rev 1_3 Commander LI90 Series Commander LI90 Series Controller User Manual Flowline Inc. 10500 Humbol

Pagina 2

10 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • DIN Rail Mounting o The LI90 series is designed to clip onto standard 35

Pagina 3

27 AUG 08 PRELIMINARY MANUAL 11 of 57 Rev 1_3 Commander LI90 Series • Dimensions Front View Side View • Temperature / Ventilation o E

Pagina 4

12 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 Electrical • Grounding Definition o Ground: The term Ground is defined as

Pagina 5

27 AUG 08 PRELIMINARY MANUAL 13 of 57 Rev 1_3 Commander LI90 Series • Wiring o Power • Connect to Earth Ground. Apply 10 - 30 VDC. o Inputs

Pagina 6

14 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 o Outputs • Commander can connect up to 6 relays, depending on the applicat

Pagina 7

27 AUG 08 PRELIMINARY MANUAL 15 of 57 Rev 1_3 Commander LI90 Series Specifications Display Graphics/Text: Yes / Yes Pixels: 128 x 64 Display Te

Pagina 8

16 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 Startup • When the Commander is powered for the first time, the LI90 series

Pagina 9

27 AUG 08 PRELIMINARY MANUAL 17 of 57 Rev 1_3 Commander LI90 Series o Application #2 / Differential Level Controller (Press F3) – Commander will

Pagina 10

18 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • If you need to change the application selection, press and hold the ESC bu

Pagina 11

27 AUG 08 PRELIMINARY MANUAL 19 of 57 Rev 1_3 Commander LI90 Series Common Information – There are several configuration aspects that are similar

Pagina 12

2 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3

Pagina 13

20 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 o For a non-scaled output. Use this description when using a transmitter th

Pagina 14

27 AUG 08 PRELIMINARY MANUAL 21 of 57 Rev 1_3 Commander LI90 Series Application #1 • Overview o Multi-Tank Display – The LI90 series in this ap

Pagina 15

22 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • General (Open Functions) o Opening Screen indicates the Level of Liquid i

Pagina 16

27 AUG 08 PRELIMINARY MANUAL 23 of 57 Rev 1_3 Commander LI90 Series • The H.O.A. screen for each relay will indicate each relay, the current sta

Pagina 17

24 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • General (Secured Functions) o The configuration of the display/input and

Pagina 18

27 AUG 08 PRELIMINARY MANUAL 25 of 57 Rev 1_3 Commander LI90 Series • Configuration of Inputs o Once in the Secured Function Section, press the

Pagina 19

26 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 Scaled 4-20 mA Output Fixed 4-20 mA Output o To change a value, first p

Pagina 20

27 AUG 08 PRELIMINARY MANUAL 27 of 57 Rev 1_3 Commander LI90 Series • Configuration of Outputs o Once in the Secured Function Section, press the

Pagina 21

28 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • Follow the sequence above: • 1. Press Enter to begin • 2. Press F1 for

Pagina 22

27 AUG 08 PRELIMINARY MANUAL 29 of 57 Rev 1_3 Commander LI90 Series Application #2 • Overview o Differential Level Controller – Commander will

Pagina 23

27 AUG 08 PRELIMINARY MANUAL 3 of 57 Rev 1_3 Commander LI90 Series Preface This manual explains how to use the Commander LI90 series controller.

Pagina 24

30 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • General (Open Functions) o Opening Screen indicates the Level of Liquid i

Pagina 25

27 AUG 08 PRELIMINARY MANUAL 31 of 57 Rev 1_3 Commander LI90 Series o In the Relay Info screen, pressing the upper right soft key will jump the s

Pagina 26

32 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • General (Secured Functions) o The configuration of the display/inputs and

Pagina 27

27 AUG 08 PRELIMINARY MANUAL 33 of 57 Rev 1_3 Commander LI90 Series • Configuration of Inputs o Once in the Secured Function Section, press the

Pagina 28

34 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 Scaled 4-20 mA Output Fixed 4-20 mA Output o To change a value, first p

Pagina 29

27 AUG 08 PRELIMINARY MANUAL 35 of 57 Rev 1_3 Commander LI90 Series • Configuration of Outputs o Once in the Secured Function Section, press the

Pagina 30

36 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • To change a value, first press the Enter key • The value will be highligh

Pagina 31

27 AUG 08 PRELIMINARY MANUAL 37 of 57 Rev 1_3 Commander LI90 Series • Relays #3 and #4 will ask for a Differential ON and Differential OFF level.

Pagina 32

38 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 Application #3 • Overview o Pump Controller – The LI90 series in this appli

Pagina 33

27 AUG 08 PRELIMINARY MANUAL 39 of 57 Rev 1_3 Commander LI90 Series • General (Open Functions) o Opening Screen indicates the Level of Liquid in

Pagina 34

4 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 Warranty Flowline warrants to the original purchaser of its products that such

Pagina 35

40 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • General (Secured Functions) o The configuration of the display, relays an

Pagina 36

27 AUG 08 PRELIMINARY MANUAL 41 of 57 Rev 1_3 Commander LI90 Series • Configuration of Inputs o Once in the Secured Function Section, press the

Pagina 37

42 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • Empty Setting – This is the display value when the tank is empty • Full S

Pagina 38

27 AUG 08 PRELIMINARY MANUAL 43 of 57 Rev 1_3 Commander LI90 Series • Volumetric Inputs o The controller will ask basic shape and dimensional in

Pagina 39

44 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 o Vertical Tanks – Select among Rectangular, Round Bottom or Cone Bottom.

Pagina 40

27 AUG 08 PRELIMINARY MANUAL 45 of 57 Rev 1_3 Commander LI90 Series o Horizontal Tanks – Select among Flat Ends or Round Ends. LengthDiameter

Pagina 41

46 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • Configuration of Outputs o Once in the Secured Function Section, press th

Pagina 42

27 AUG 08 PRELIMINARY MANUAL 47 of 57 Rev 1_3 Commander LI90 Series o To change a value, first press the Enter key • The value will be highlight

Pagina 43

48 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 o Select either an Independent Relay #5 and #6 or Duplexing Relays #5 and #6

Pagina 44

27 AUG 08 PRELIMINARY MANUAL 49 of 57 Rev 1_3 Commander LI90 Series • To use as Independent Relay mode, press the IND soft key and then press Nex

Pagina 45 - Diameter

27 AUG 08 PRELIMINARY MANUAL 5 of 57 Rev 1_3 Commander LI90 Series Table of Contents Safety Warnings and Guidelines ...

Pagina 46

50 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • To change to an Alternation mode, press the DUP soft key and then press Ne

Pagina 47

27 AUG 08 PRELIMINARY MANUAL 51 of 57 Rev 1_3 Commander LI90 Series Appendix Simulation Mode This mode enables the end user to simulate changes i

Pagina 48

52 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 o Next, press the Begin Simulation soft key to proceed or the Back soft key

Pagina 49

27 AUG 08 PRELIMINARY MANUAL 53 of 57 Rev 1_3 Commander LI90 Series Factory Reset • At anytime, the Commander Series can be reset to its original

Pagina 50

54 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 Changing Applications • If you need to change the application selection, pr

Pagina 51

27 AUG 08 PRELIMINARY MANUAL 55 of 57 Rev 1_3 Commander LI90 Series Removable Media for Data Transfer Overview • All Commander models provide a

Pagina 52

56 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 Using Removable Media to Log Data • The Commander has the feature of loggin

Pagina 53

27 AUG 08 PRELIMINARY MANUAL 57 of 57 Rev 1_3 Commander LI90 Series Troubleshooting PROBLEM SOLUTION No power to the transmitters The Analog

Pagina 54

6 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 Safety Warnings and Guidelines When found on the product, the following symbo

Pagina 55

27 AUG 08 PRELIMINARY MANUAL 7 of 57 Rev 1_3 Commander LI90 Series Introduction The Commander Series has three unique application configurations.

Pagina 56 - Empty and Full settings)

8 of 57 PRELIMINARY MANUAL 27 AUG 08 Commander LI90 Series Rev 1_3 • Pump Controller (Application #3) – Commander will accept a single 4-20 mA

Pagina 57

27 AUG 08 PRELIMINARY MANUAL 9 of 57 Rev 1_3 Commander LI90 Series Installation • Overview o The mechanical installation greatly affects the op